Maintenance and inspection of medical operation lights

Dec 07, 2024 Leave a message

‌Regular inspection‌: Check the tightness of the rotating body chassis connection nuts and other connection screws at least once a month to prevent loosening‌.
‌Power line inspection‌: Check the power line at least once a quarter to confirm that there is no oxidation on the connection terminals, and use a megohmmeter to check whether the connection of the protective grounding wire is secure‌.
‌Professional maintenance‌: Regularly invite professional technicians to conduct comprehensive inspections and maintenance of the shadowless lamp to ensure that the equipment is always in good working condition‌.
‌Shutdown procedure‌: Observe that the shadowless lamp is completely extinguished and all indicator lights on the control panel are extinguished before leaving the operating room‌.